Principle:
The test uses an anti-human cTnl monoclonal antibody conjugated with fluorescence latex and another anti-human cTnl monoclonal antibody coated on the test line. After the sample has been applied to the test strip, the fluorescence latex-labelled anti-human cTni monoclonal antibody binds with the cTnl in sample and forms a marked antigen-anlibody complex. This complex moves to the test delection zone by capillary action. Then marked antigen-antibody complex is captured on the test line by the anti-human cTnl monoclonal antibody. The fluorescence intensity of the test line increases in proportion to the amount of cTnl in sample.
Then, the test card is inserted into ACCUDx CQ Immunofluorescence Quantitative Analyzer. The concentration of cTnl in sample will be measured and displayed on the screen. The value will be stored in ACCUDx CQ Analyzer. The result can be transmitted to the laboratory or hospital information system.